Wattigney W A, Webber L S, Lawrence M D, Berenson G S
Department of Applied Health Sciences, Tulane School of Public Health and Tropical Medicine, New Orleans, LA 70112-2824, USA.
Am J Hypertens. 1996 Mar;9(3):256-62. doi: 10.1016/0895-7061(95)00334-7.
The Dinamap 845XT automatic blood pressure monitor, Dinamap 8100 (an update model), and two mercury sphygmomanometers were compared in 417 school-aged children examined in the spring of 1992 as part of the Bogalusa Heart Study. This study was conducted in the nearby community of Franklinton, Louisiana, to verify data obtained as part of a cross-sectional survey (1987 to 1988) of school-aged children in Bogalusa. Systolic blood pressure levels were on the average 3 mm Hg higher on the Dinamap instruments than on the sphygmomanometers. Mean levels of diastolic blood pressure using either Dinamap instrument were slightly higher until eight years of age and then were considerably lower than mercury sphygmomanometer fourth phase readings. Diastolic blood pressure levels on the Dinamap 8100 were 4 mm Hg lower than on the Dinamap 845XT. Height was identified as the predominant predictor variable of differences in diastolic blood pressure between the mercury sphygmomanometer and either Dinamap instrument. A 10% random sample of children was reexamined each screening day in the cross-sectional survey to estimate measurement errors. The diastolic readings of the Dinamap 845XT had a lower intraclass correlation (0.68) compared to the mercury sphygmomanometers (0.83 fourth phase and 0.76 fifth phase). The Dinamap offers the ease of measuring systolic blood pressure although the diastolic blood pressure appears to be biased and especially low, particularly on the new 8100 model.
作为博加卢萨心脏研究的一部分,1992年春季对417名学龄儿童进行了研究,比较了Dinamap 845XT自动血压监测仪、Dinamap 8100(升级版)和两台汞柱式血压计。这项研究在路易斯安那州弗兰克林顿附近的社区进行,以核实作为博加卢萨学龄儿童横断面调查(1987年至1988年)一部分所获得的数据。Dinamap仪器测量的收缩压水平平均比血压计高3毫米汞柱。使用任一台Dinamap仪器测量的舒张压平均水平在8岁之前略高,之后则明显低于汞柱式血压计的第四阶段读数。Dinamap 8100上的舒张压水平比Dinamap 845XT低4毫米汞柱。身高被确定为汞柱式血压计与任一台Dinamap仪器之间舒张压差异的主要预测变量。在横断面调查中,每个筛查日对10%的儿童随机样本进行重新检查,以估计测量误差。与汞柱式血压计(第四阶段为0.83,第五阶段为0.76)相比,Dinamap 845XT的舒张压读数组内相关性较低(0.68)。Dinamap测量收缩压较为简便,尽管舒张压似乎存在偏差且尤其偏低,特别是在新的8100型号上。