Miller J M, Rhoden D L
Centers for Disease Control, Atlanta, Georgia 30333.
J Clin Microbiol. 1991 Jun;29(6):1143-7. doi: 10.1128/jcm.29.6.1143-1147.1991.
The Biolog Identification System (Biolog, Inc., Hayward, Calif.) is a new bacterial identification method that establishes an identification based on the exchange of electrons generated during respiration, leading to a subsequent tetrazolium-based color change. This system tests the ability of a microorganism to oxidize a panel of 95 different carbon sources. We report on a preliminary investigation of the ability of the instrument to identify, using its computer-driven enzyme immunoassay reader, a diverse group of clinically relevant members of the family Enterobacteriaceae and gram-negative non-Enterobacteriaceae. The Biolog reported identifications (correct or incorrect) for 266 of 352 organisms tested (75.6%). Of the 266 identifications reported, 87.3% were correct at the genus level and 75.6% were correct at the species level at 24 h. In the total study of 352 strains, 46.6% were correct to the species level at 4 h and 57.1% were correct to the species level at 24 h. The error rate was 10.4% after 4 h and 9.6% after 24 h. The Biolog performed well with many genera, but problems were encountered with some strains of Klebsiella, Enterobacter, and Serratia. We found the system to be versatile and easy to use.
Biolog鉴定系统(Biolog公司,加利福尼亚州海沃德)是一种新的细菌鉴定方法,它基于呼吸过程中产生的电子交换来进行鉴定,进而导致随后基于四氮唑的颜色变化。该系统测试微生物氧化一组95种不同碳源的能力。我们报告了一项初步调查,研究该仪器利用其计算机驱动的酶免疫分析读数器鉴定肠杆菌科和革兰氏阴性非肠杆菌科的一组不同临床相关菌属的能力。在测试的352种微生物中,Biolog报告了266种的鉴定结果(正确或错误)(75.6%)。在报告的266种鉴定结果中,24小时时属水平的正确率为87.3%,种水平的正确率为75.6%。在对352株菌株的总体研究中,4小时时种水平的正确率为46.6%,24小时时种水平的正确率为57.1%。4小时后的错误率为10.4%,24小时后的错误率为9.6%。Biolog对许多菌属的鉴定效果良好,但在一些克雷伯菌属、肠杆菌属和沙雷菌属菌株中遇到了问题。我们发现该系统用途广泛且易于使用。