Al-Farhan Haya M, Al-Otaibi Wafa'a Majed
Department of Optometry and Vision Sciences, College of Applied Medicine Science, King Saud University, Riyadh, Kingdom of Saudia Arabia.
Clin Ophthalmol. 2012;6:1037-43. doi: 10.2147/OPTH.S32955. Epub 2012 Jul 6.
To compare the precision of central corneal thickness (CCT) measurements taken with the handheld ultrasound pachymeter (USP), ultrasound biomicroscopy (UBM), and the Artemis-2 very high frequency ultrasound scanner (VHFUS) on normal subjects.
Prospective study.
One eye from each of 61 normal subjects was randomly selected for this study. The measurements of the CCT were taken with the USP, VHFUS, and UBM. Results were compared statistically using repeated-measures analysis of variance (ANOVA), Pearson's correlation coefficient, and limits of agreement.
The average CCT (± standard deviation) was 530.1 ± 30.5 μm, 554.9 ± 31.7 μm, and 559.5 ± 30.7 μm for UBM, VHFUS, and USP respectively. The intraobserver repeatability analyses of variance are not significant for USP, UBM, and VHFUS. P-values were 0.17, 0.19, and 0.37 respectively. Repeated-measures ANOVA showed a significant difference between the three different methods of measuring CCT (P = 0.0001). The ANOVA test revealed no statistically significant difference between USP and VHFUS (P > 0.05), yet statistical significant differences with UBM versus USP and UBM versus VHFUS (P < 0.001). There were high correlations between the three instruments (P < 0.0001). The mean differences (and upper/lower limits of agreement) for CCT measurements were 29.4 ± 14.3 (2.7/56), 4.6 ± 8.6 (-14.7/23.8), and -24.8 ± 13.1 (-50.4/0.8) for USP versus UBM, USP versus VHFUS, and UBM versus VHFUS, respectively.
The UBM produces CCT measurements that vary significantly from those returned by the USP and the VHFUS, suggesting that the UBM may not be used interchangeably with either equipment for monitoring the CCT in the clinical setting.
比较手持超声角膜测厚仪(USP)、超声生物显微镜(UBM)和Artemis - 2甚高频超声扫描仪(VHFUS)对正常受试者中央角膜厚度(CCT)测量的精度。
前瞻性研究。
从61名正常受试者中随机选取一只眼睛进行本研究。使用USP、VHFUS和UBM测量CCT。采用重复测量方差分析(ANOVA)、Pearson相关系数和一致性界限对结果进行统计学比较。
UBM、VHFUS和USP测量的平均CCT(±标准差)分别为530.1±30.5μm、554.9±31.7μm和559.5±30.7μm。USP、UBM和VHFUS的观察者内重复性方差分析无统计学意义。P值分别为0.17、0.19和0.37。重复测量ANOVA显示三种不同测量CCT的方法之间存在显著差异(P = 0.0001)。ANOVA检验显示USP和VHFUS之间无统计学显著差异(P>0.05),但UBM与USP以及UBM与VHFUS之间存在统计学显著差异(P<0.001)。三种仪器之间存在高度相关性(P<0.0001)。USP与UBM、USP与VHFUS、UBM与VHFUS测量CCT的平均差异(以及一致性上限/下限)分别为29.4±14.3(2.7/56)、4.6±8.6(-14.7/23.8)和-24.8±13.1(-50.4/0.8)。
UBM测量的CCT与USP和VHFUS测量的结果有显著差异,这表明在临床环境中监测CCT时,UBM可能不能与这两种设备互换使用。