Key Laboratory of Vision Science, School of Optometry and Ophthalmology and Eye Hospital, Wenzhou Medical College, Ministry of Health PR China.
J Cataract Refract Surg. 2011 Feb;37(2):341-8. doi: 10.1016/j.jcrs.2010.08.044.
To compare central corneal thickness (CCT), anterior chamber depth (ACD), and keratometry (K) readings measured using optical low-coherence reflectometry (OLCR) biometry and high-resolution rotating Scheimpflug photography.
Eye Hospital of Wenzhou Medical College, Wenzhou, China.
Comparative case series.
The CCT, ACD endothelium to lens, ACD epithelium to lens, and K (mean; in flattest meridian; in steepest meridian) were measured 5 times using the LenStar/Biograph OLCR biometer and 3 times with the Pentacam Scheimpflug system in eyes of healthy volunteers. Concordance was evaluated using paired t tests, the Pearson correlation, and Bland-Altman analyses.
The CCT, ACD endothelium to lens, and ACD epithelium to lens measured with the Scheimpflug system were slightly, albeit significantly, higher than with the OLCR biometer (P<.05); the respective 95% limits of agreement (LoA) were -8.2 μm to 15.7 μm, -0.11 to 0.15 mm, and -0.13 to 0.17 mm. However, the Scheimpflug system gave significantly flatter readings for K in the flattest meridian (95% LoA, -0.54 to 0.32 diopters [D]), K in the steepest meridian (95% LoA, -0.63 to 0.45 D), and mean K (95% LoA, -0.53 to 0.33 D) (P<.001). The CCT, ACD, and K readings were all highly correlated between the 2 devices (r >0.95, P<.001).
The CCT and ACD measurements with the OLCR biometer and Scheimpflug system can be used interchangeably in healthy young subjects. However, for K measurements, these devices have wide LoA so may not be interchangeable under certain clinical circumstances.
比较光学低相干反射计(OLCR)生物测量和高分辨率旋转 Scheimpflug 摄影测量的中央角膜厚度(CCT)、前房深度(ACD)和角膜曲率(K)读数。
中国温州医科大学眼视光医院。
对比病例系列。
使用 LenStar/Biograph OLCR 生物测量仪 5 次和 Pentacam Scheimpflug 系统 3 次测量健康志愿者眼的 CCT、内皮至晶状体的 ACD、上皮至晶状体的 ACD 和 K(平均值;最平坦子午线;最陡峭子午线)。使用配对 t 检验、Pearson 相关和 Bland-Altman 分析评估一致性。
Scheimpflug 系统测量的 CCT、内皮至晶状体的 ACD 和上皮至晶状体的 ACD 略高,但差异有统计学意义(P<.05);各自的 95%一致性界限(LoA)分别为-8.2μm 至 15.7μm、-0.11 至 0.15mm 和-0.13 至 0.17mm。然而,Scheimpflug 系统在最平坦子午线的 K 读数明显较平坦(95%LoA,-0.54 至 0.32 屈光度[D])、最陡峭子午线的 K 读数(95%LoA,-0.63 至 0.45 D)和平均 K 读数(95%LoA,-0.53 至 0.33 D)(P<.001)。两种设备之间的 CCT、ACD 和 K 读数均高度相关(r>0.95,P<.001)。
在健康年轻受试者中,OLCR 生物测量仪和 Scheimpflug 系统的 CCT 和 ACD 测量可以互换使用。然而,对于 K 测量,这些设备的 LoA 较宽,因此在某些临床情况下可能无法互换。