Xiao-Hong Zhou, Lan-Hua Liu, Wei-Qi Xu, Bao-Dong Song, Jian-Wu Sheng, Miao He, Han-Chang Shi
1] State Key Joint Laboratory of ESPC, School of Environment, Tsinghua University, Beijing 10084, China [2] Collaborative Innovation Center for Advanced Water Pollution Control Technology and Equipment, Nanjing, China.
State Key Joint Laboratory of ESPC, School of Environment, Tsinghua University, Beijing 10084, China.
Sci Rep. 2014 Apr 4;4:4572. doi: 10.1038/srep04572.
This paper proposed a compact and portable planar waveguide evanescent wave immunosensor (EWI) for highly sensitive detection of BPA. The incident light is coupled into the planar waveguide chip via a beveled angle through undergoing total internal reflection, where the evanescent wave field forms and excites the binding fluorophore-tagged antibodies on the chip surface. Typical calibration curves obtained for BPA has detection limits of 0.03 μg/L. Linear response for BPA ranged from 0.124 μg/L-9.60 μg/L with 50% inhibition concentration for BPA of 1.09 ± 0.25 μg/L. The regeneration of the planar optical waveguide chip allows the performance of more than 300 assay cycles within an analysis time of about 20 min for each assay cycle. By application of effective pretreatment procedure, the recoveries of BPA in real water samples gave values from 88.3% ± 8.5% to 103.7% ± 3.5%, confirming its application potential in the measurement of BPA in reality.
本文提出了一种用于高灵敏度检测双酚A(BPA)的紧凑型便携式平面波导倏逝波免疫传感器(EWI)。入射光通过斜角以全内反射的方式耦合到平面波导芯片中,在那里形成倏逝波场并激发芯片表面结合的荧光团标记抗体。获得的BPA典型校准曲线的检测限为0.03μg/L。BPA的线性响应范围为0.124μg/L至9.60μg/L,BPA的50%抑制浓度为1.09±0.25μg/L。平面光波导芯片的再生使得在每个检测周期约20分钟的分析时间内能够进行超过300次检测循环。通过应用有效的预处理程序,实际水样中BPA的回收率为88.3%±8.5%至103.7%±3.5%,证实了其在实际测量BPA中的应用潜力。