Maeda Naoyuki, Ueki Ryotaro, Fuchihata Mutsumi, Fujimoto Hisataka, Koh Shizuka, Nishida Kohji
Department of Ophthalmology, Osaka University Graduate School of Medicine, Room E7, 2-2 Yamadaoka, Suita, Osaka, 565-0871, Japan,
Jpn J Ophthalmol. 2014 Nov;58(6):483-9. doi: 10.1007/s10384-014-0344-2. Epub 2014 Sep 5.
To compare the corneal biomechanical properties of eyes that have undergone penetrating keratoplasty (PK), deep anterior lamellar keratoplasty (DALK), and Descemet stripping automated endothelial keratoplasty (DSAEK).
This case-control study comprised 20 post-PK eyes, 14 post-DALK eyes, 15 post-DSAEK eyes, and 50 normal control eyes. A dynamic Scheimpflug analyzer (the Corvis ST) was used to evaluate the corneal biomechanical properties including deformation amplitude (DA) and radius at the highest concavity (R hc).
In post-PK eyes, the mean DA was 1.20 ± 0.13 mm, which was significantly higher than those of the control eyes (1.07 ± 0.09) and the post-DSAEK eyes (1.08 ± 0.12). The DA (1.18 ± 0.18) in the post-DALK eyes was significantly higher than in the control eyes. The R hc in the post-PK (6.34 ± 0.37 mm), -DALK (6.04 ± 1.22), and -DSAEK (6.44 ± 0.58) eyes was significantly smaller than in the control eyes (7.57 ± 0.78).
The dynamic Scheimpflug analyzer provides a method to obtain new biomechanical information on the cornea such as the DA and R hc, and these parameters differed among eyes that had undergone 3 different types of corneal surgery. Abnormalities in these parameters after the different corneal transplantation techniques may indicate larger deviations in the stress-strain reaction of the cornea and more uncertainty in the intraocular pressure measurements than in normal eyes.
比较穿透性角膜移植术(PK)、深板层角膜移植术(DALK)和Descemet膜剥脱自动内皮角膜移植术(DSAEK)术后眼睛的角膜生物力学特性。
本病例对照研究包括20只PK术后眼、14只DALK术后眼、15只DSAEK术后眼和50只正常对照眼。使用动态Scheimpflug分析仪(Corvis ST)评估角膜生物力学特性,包括变形幅度(DA)和最高凹度处的半径(R hc)。
PK术后眼中,平均DA为1.20±0.13mm,显著高于对照眼(1.07±0.09)和DSAEK术后眼(1.08±0.12)。DALK术后眼的DA(1.18±0.18)显著高于对照眼。PK术后眼(6.34±0.37mm)、DALK术后眼(6.04±1.22)和DSAEK术后眼(6.44±0.58)的R hc显著小于对照眼(7.57±0.78)。
动态Scheimpflug分析仪提供了一种获取角膜新生物力学信息(如DA和R hc)的方法,这些参数在接受3种不同类型角膜手术的眼睛中有所不同。不同角膜移植技术后这些参数的异常可能表明角膜应力应变反应的偏差更大,与正常眼睛相比,眼压测量的不确定性也更大。