同轴2.2毫米微小切口与标准3.0毫米超声乳化术后角膜生物力学特性的变化

Corneal biomechanical properties changes after coaxial 2.2-mm microincision and standard 3.0-mm phacoemulsification.

作者信息

Zhang Zhe, Yu Hua, Dong Hui, Wang Li, Jia Ya-Ding, Zhang Su-Hua

机构信息

Clinical College of Ophthalmology, Tianjin Medical University, Tianjin Eye Hospital, Tianjin Key Laboratory of Ophthalmology and Visual Science, Tianjin Eye Institute, Tianjin 300070, China; Shanxi Eye Hospital, Taiyuan 030002, Shanxi Province, China.

Shanxi Eye Hospital, Taiyuan 030002, Shanxi Province, China.

出版信息

Int J Ophthalmol. 2016 Feb 18;9(2):230-4. doi: 10.18240/ijo.2016.02.08. eCollection 2016.

Abstract

AIM

To compare the changes in corneal biomechanics measured by ocular response analyzer (ORA) after 2.2-mm microincision cataract surgery and 3.0-mm standard coaxial phacoemulsification.

METHODS

The prospective nonrandomized study comprised eyes with cataract that had 2.2-mm coaxial microincision or 3.0-mm standard incision phacoemulsification. The corneal hysteresis (CH), corneal resistance factor (CRF), corneal-compensated intraocular pressure (IOPcc) and Goldmann-correlated intraocular pressure (IOPg) were measured by ORA preoperatively and at 1d, 1-, 2-, 3- and 4-week postoperatively. Results were analyzed and compared between groups.

RESULTS

In both groups, CH decreased in the immediate postoperative period (P<0.05), returned to the preoperative level at one week (P=0.249) in the 2.2-mm group, and at two weeks in the 3.0-mm group (P=0.264); there was no significant change in CRF values. In 2.2-mm group, mean IOPcc and IOPg increased at 1d postoperatively (both P<0.05), and returned to preoperative level at one week (P=0.491 and P=0.923, respectively). In 3.0-mm group, mean IOPcc and IOPg increased at 1d and 1wk postoperatively (P=0.005 and P=0.029, respectively), and returned to preoperative level at 2wk (P=0.347 and P=0.887, respectively).

CONCLUSION

Significant differences between preoperative and postoperative corneal biomechanical values were found for CH, IOPcc and IOPg. But the recovery time courses were different between the two groups. The 2.2-mm coaxial microincision cataract surgery group seemed recovery faster compared to the 3.0-mm standard coaxial phacoemulsification group.

摘要

目的

比较2.2毫米微小切口白内障手术和3.0毫米标准同轴超声乳化术后用眼反应分析仪(ORA)测量的角膜生物力学变化。

方法

这项前瞻性非随机研究纳入了接受2.2毫米同轴微小切口或3.0毫米标准切口超声乳化白内障手术的白内障患者。术前以及术后1天、1周、2周、3周和4周用ORA测量角膜滞后(CH)、角膜阻力因子(CRF)、角膜补偿眼压(IOPcc)和Goldmann相关眼压(IOPg)。对两组结果进行分析和比较。

结果

两组患者术后即刻CH均降低(P<0.05),2.2毫米组在术后1周恢复到术前水平(P=0.249),3.0毫米组在术后2周恢复到术前水平(P=0.264);CRF值无显著变化。2.2毫米组术后1天平均IOPcc和IOPg升高(均P<0.05),1周时恢复到术前水平(分别为P=0.491和P=0.923)。3.0毫米组术后1天和1周平均IOPcc和IOPg升高(分别为P=0.005和P=0.029),2周时恢复到术前水平(分别为P=0.347和P=0.887)。

结论

CH、IOPcc和IOPg的术前和术后角膜生物力学值存在显著差异。但两组的恢复时间进程不同。与3.0毫米标准同轴超声乳化组相比,2.2毫米同轴微小切口白内障手术组似乎恢复得更快。

相似文献

10
Corneal Biomechanical Characteristics in Patıents with Behçet Disease.白塞病患者的角膜生物力学特征
Semin Ophthalmol. 2016;31(5):439-45. doi: 10.3109/08820538.2014.962168. Epub 2014 Nov 13.

引用本文的文献

10
The Effect of Strabismus Muscle Surgery on Corneal Biomechanics.斜视肌肉手术对角膜生物力学的影响。
J Ophthalmol. 2018 Sep 16;2018:8072140. doi: 10.1155/2018/8072140. eCollection 2018.

本文引用的文献

9
Biomechanical properties of the cornea in Fuchs' corneal dystrophy.富克斯角膜营养不良中角膜的生物力学特性。
Invest Ophthalmol Vis Sci. 2009 Jul;50(7):3199-202. doi: 10.1167/iovs.08-3312. Epub 2009 Feb 28.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验