Backes B A, Cavalieri S J, Rudrik J T, Britt E M
J Clin Microbiol. 1984 Jun;19(6):744-7. doi: 10.1128/jcm.19.6.744-747.1984.
Comparisons of the antimicrobial susceptibility patterns of clinical isolates of the family Enterobacteriaceae and Pseudomonas aeruginosa were made by using Kirby-Bauer disk diffusion, the MicroMedia system broth microdilution, and the AutoMicrobic system. The Kirby-Bauer system compared favorably with the MicroMedia system (95% essential correlation with 90% complete agreement). The AutoMicrobic system also compared favorably with both the Kirby-Bauer system (93% essential correlation with 87% complete agreement) and the MicroMedia system (97% essential correlation with 85% complete agreement). These data indicate that the AutoMicrobic system results are comparable to those of the MicroMedia and the Kirby-Bauer systems. Furthermore, the AutoMicrobic system offers a more rapid system (6 to 10 h), than the traditional systems, without sacrificing accuracy (16 to 24 h).
采用 Kirby-Bauer 纸片扩散法、MicroMedia 系统肉汤微量稀释法和自动微生物鉴定系统,对肠杆菌科临床分离株和铜绿假单胞菌的抗菌药敏模式进行了比较。Kirby-Bauer 系统与 MicroMedia 系统相比表现良好(95%基本相关性,90%完全一致性)。自动微生物鉴定系统与 Kirby-Bauer 系统(93%基本相关性,87%完全一致性)和 MicroMedia 系统(97%基本相关性,85%完全一致性)相比也表现良好。这些数据表明,自动微生物鉴定系统的结果与 MicroMedia 系统和 Kirby-Bauer 系统的结果相当。此外,自动微生物鉴定系统比传统系统提供了更快的检测速度(6 至 10 小时),且不牺牲准确性(16 至 24 小时)。