Validation of indirect calorimetry for measurement of energy expenditure in healthy volunteers undergoing pressure controlled non-invasive ventilation support.

机构信息

Department of Anaesthesiology, Intensive Care, Emergency Care and Pain Medicine, Turku University Hospital and University of Turku, Kiinamyllynkatu 4-8, 20521, Turku, Finland.

出版信息

J Clin Monit Comput. 2012 Feb;26(1):37-43. doi: 10.1007/s10877-011-9331-z. Epub 2011 Dec 30.

Abstract

The aim of this validation study was to assess the reliability of gas exchange measurement with indirect calorimetry among subjects who undergo non-invasive ventilation (NIV). Oxygen consumption (VO2) and carbon dioxide production (VCO2) were measured in twelve healthy volunteers. Respiratory quotient (RQ) and resting energy expenditure (REE) were then calculated from the measured VO2 and VCO2 values. During the measurement period the subjects were breathing spontaneously and ventilated using NIV. Two different sampling air flow values 40 and 80 l/min were used. The gas leakage from the measurement setup was assessed with a separate capnograph. The mean weight of the subjects was 93 kg. Their mean body mass index was 29 (range 22-40) kg/m2. There was no statistically significant difference in the measured values for VO2, VCO2, RQ and REE during NIV-supported breathing and spontaneous breathing. The change of sampling air flow had no statistically significant effect on any of the above parameters. We found that REE can be accurately measured with an indirect calorimeter also during NIV-supported breathing and the change of sampling air flow does not distort the gas exchange measurement. A higher sampling air flow in indirect calorimetry decreases the possibility for air leakages in the measurement system and increases the reliability of REE measurement.

摘要

本验证研究旨在评估在接受无创通气(NIV)的受试者中,间接测热法测量气体交换的可靠性。我们测量了 12 名健康志愿者的氧气消耗(VO2)和二氧化碳产生量(VCO2)。然后,根据测量的 VO2 和 VCO2 值计算呼吸商(RQ)和静息能量消耗(REE)。在测量期间,受试者进行自主呼吸并使用 NIV 进行通气。使用了两种不同的采样空气流量值 40 和 80 l/min。使用单独的二氧化碳描记器评估测量装置的气体泄漏情况。受试者的平均体重为 93 公斤。他们的平均体重指数为 29(范围 22-40)kg/m2。在 NIV 支持呼吸和自主呼吸期间,VO2、VCO2、RQ 和 REE 的测量值之间没有统计学上的显著差异。采样空气流量的变化对上述任何参数均无统计学显著影响。我们发现,间接测热法也可以在 NIV 支持呼吸期间准确测量 REE,并且采样空气流量的变化不会扭曲气体交换测量。间接测热法中较高的采样空气流量降低了测量系统中空气泄漏的可能性,并提高了 REE 测量的可靠性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验