Wanichwecharungruang Boonsong, Amornpetchsathaporn Anyarak, Wongwijitsook Wisakorn, Kongsomboon Kittipong, Chantra Somporn
Department of Ophthalmology, Rajavithi Hospital and Rangsit Medical College, Bangkok, Thailand.
Department of Ophthalmology, Priest Hospital, Bangkok, Thailand.
PLoS One. 2022 Mar 21;17(3):e0265844. doi: 10.1371/journal.pone.0265844. eCollection 2022.
To investigate agreement between 2 swept source OCT biometers, IOL Master700 and Anterion, in various ocular biometry and intraocular lens (IOL) calculations of primary angle-closure disease (PACD).
Rajavithi Hospital, Bangkok, Thailand.
Prospective comparative study.
This study conducted in a tertiary eye care center involving biometric measurements obtained with 2 devices in phakic eye with diagnosis of PACD. Mean difference and intraclass correlation coefficient (ICC) with confidence limits were assessed, and calculations of estimated residual refraction of the IOL were analysed using Barrett's formula.
Sixty-nine eyes from 45 PACD patients were enrolled for the study. Excellent agreement of various parameters was revealed, with ICC (confidence limits) of K1 = 0.953 (0.861-0.979), K2 = 0.950 (0.778-0.98), ACD = 0.932 (0.529-0.978), WTW = 0.775 (0.477-0.888), and LT = 0.947 (0.905-0.97). Mean difference of axial length (AL) was -0.01 ± 0.02 mm with ICC of 1.000. IOL calculation was assessed with Barrett's formula, and Bland-Altman plot showed excellent agreement in the results of the 2 devices for the IOL power and estimated post-operative residual refraction (EPR).
Mean differences of biometric parameters, obtained with IOL Master700 and Anterion, were small, and ICC showed excellent concordance. No clinical relevance in calculation of IOL power was found, and the two devices appeared to be comparably effective in clinical practice.
研究两种扫频源光学相干断层扫描生物测量仪(IOL Master700和Anterion)在原发性闭角型青光眼(PACD)的各种眼生物测量和人工晶状体(IOL)计算中的一致性。
泰国曼谷拉贾维蒂医院。
前瞻性比较研究。
本研究在一家三级眼科护理中心进行,对诊断为PACD的有晶状体眼使用两种设备进行生物测量。评估平均差异和具有置信限的组内相关系数(ICC),并使用巴雷特公式分析IOL估计残余屈光度的计算。
45例PACD患者的69只眼睛纳入研究。结果显示各种参数具有良好的一致性,K1的ICC(置信限)为0.953(0.861 - 0.979),K2为0.950(0.778 - 0.98),前房深度(ACD)为0.932(0.529 - 0.978)横向白到白距离(WTW)为0.775(0.477 - 0.888),晶状体厚度(LT)为0.947(0.905 - 0.97)。眼轴长度(AL)的平均差异为-0.01±0.02mm,ICC为1.000。使用巴雷特公式评估IOL计算,Bland-Altman图显示两种设备在IOL屈光度和估计术后残余屈光度(EPR)结果方面具有良好的一致性。
IOL Master700和Anterion获得的生物测量参数平均差异较小,ICC显示出良好的一致性。在IOL屈光度计算中未发现临床相关性,这两种设备在临床实践中似乎具有相当的有效性。