Passi Samuel F, Thompson Atalie C, Gupta Preeya K
Department of Ophthalmology, Duke University Eye Center, Durham, NC, USA,
Clin Ophthalmol. 2018 Nov 1;12:2245-2251. doi: 10.2147/OPTH.S182898. eCollection 2018.
To compare the agreement and efficiency of a swept source-optical coherence tomography biometer, IOLMaster 700 (IOLM700), and a low-coherence optical reflectometry biometer, LENSTAR LS 900 (LS900), when acquiring biometry measurements during cataract evaluation.
A retrospective chart review of biometry measurements that were performed in 64 eyes of 32 patients on the same day using the LS900 and the IOLM700. The total image acquisition time per subject was compared between the two machines using a Wilcoxon rank-sum test. Bland-Altman plots showing the mean difference and 95% limits of agreement were graphed. Intraclass correlation coefficients (ICCs) were calculated for the mean axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and two keratometry mean values (K1 and K2) that were output from each device.
The average time to complete biometry measurements in both eyes was significantly shorter for the IOLM700 compared with the LS900 (44.5±12.4 vs 168.8±67.2 seconds, <0.001). The Bland-Altman analysis and ICCs showed high degrees of agreement for the mean biometry values (ICC: AL 0.9999, ACD 0.9993, LT 0.9571, K1 0.9922, K2 0.9926) generated by the two devices.
There was a high level of agreement between the mean biometry output measures for IOLM700 and LS900. However, it took ~73% less time on average to acquire the images when using the IOLM700 compared with the LS900. In a busy clinic setting, use of the IOLM700 for biometry measures may save time and prove more efficient.
比较在白内障评估过程中使用扫频源光学相干断层扫描生物测量仪IOLMaster 700(IOLM700)和低相干光学反射测量生物测量仪LENSTAR LS 900(LS900)进行生物测量时的一致性和效率。
对32例患者的64只眼在同一天使用LS900和IOLM700进行生物测量的回顾性病历分析。使用Wilcoxon秩和检验比较两台仪器每位受试者的总图像采集时间。绘制Bland-Altman图,显示平均差异和95%一致性界限。计算两台设备输出的平均眼轴长度(AL)、前房深度(ACD)、晶状体厚度(LT)以及两个角膜曲率平均值(K1和K2)的组内相关系数(ICC)。
与LS900相比,IOLM700测量双眼生物测量的平均完成时间显著更短(44.5±12.4秒对168.8±67.2秒,<0.001)。Bland-Altman分析和ICC显示两台设备生成的平均生物测量值具有高度一致性(ICC:AL 0.9999,ACD 0.9993,LT 0.9571,K1 0.9922,K2 0.9926)。
IOLM700和LS900的平均生物测量输出指标之间具有高度一致性。然而,与LS900相比,使用IOLM700采集图像的平均时间减少了约73%。在繁忙的临床环境中,使用IOLM700进行生物测量可能节省时间并证明更高效。