Center for Ophthalmic Optics and Lasers, Doheny Eye Institute and Department of Ophthalmology, Keck School of Medicine, University of Southern California, Los Angeles, California 90033, USA.
J Cataract Refract Surg. 2010 May;36(5):826-31. doi: 10.1016/j.jcrs.2009.11.016.
To evaluate the repeatability of Fourier-domain optical coherence tomography (OCT) pachymetric mapping and compare central corneal thickness (CCT) measurements by OCT, ultrasound pachymetry, and scanning-slit tomography.
Doheny Eye Institute, University of Southern California, Los Angeles, California, USA.
A Fourier-domain OCT system was used to map the corneal thickness in normal eyes with scans centered on the corneal vertex or the pupil. Repeatability of central and pericentral map sectors was assessed by pooled standard deviation. The CCT measurements were compared between the OCT, ultrasound, and scanning-slit devices.
Pupil centration (SD: 1.3 microm central, 1.8 to 3.8 microm pericentral) provided better repeatability than vertex centration (SD: 1.7 microm central, 2.4 to 5.7 microm pericentral) in all sectors (P<.035). The mean CCT was 536.9 microm +/- 27.0 (SD) by OCT, 556.6 +/- 30.5 microm by ultrasound, and 537.2 +/- 32.6 microm by scanning-slit tomography (acoustic factor 0.92). The CCT measured by OCT was significantly thinner than by ultrasound pachymetry (P = .007; mean difference -19.7 microm; 95% limits of agreement [LoA], -40.4 to 0.9 microm) but not than by scanning-slit tomography (P = .2637; mean difference -0.3 microm; 95% LoA, -24.0 to 23.5 microm). The CCT by OCT correlated well with ultrasound and scanning-slit CCTs (r = 0.940 and r = 0.934, respectively).
Pachymetric mapping with Fourier-domain OCT was highly repeatable. Repeatability was better with pupil-centered scans than with corneal vertex-centered scans. Ultrasound pachymetry, Fourier-domain OCT, and scanning-slit tomography should not be used interchangeably for CCT assessment.
评估傅里叶域光学相干断层扫描(OCT)角膜厚度测绘的可重复性,并比较 OCT、超声角膜测厚仪和扫描 slit 体层摄影术测量中央角膜厚度(CCT)的结果。
美国加利福尼亚州洛杉矶南加州大学 Doheny 眼科研究所。
傅里叶域 OCT 系统用于对以角膜顶点或瞳孔为中心的正常眼进行角膜厚度测绘。通过汇总标准差评估中央和周边测绘区的重复性。比较 OCT、超声和扫描 slit 设备的 CCT 测量值。
与角膜顶点定位(中央 1.7μm,周边 2.4 到 5.7μm;SD)相比,瞳孔定位(中央 1.3μm,周边 1.8 到 3.8μm;SD)在所有区均能提供更好的可重复性(P<.035)。OCT 测量的 CCT 平均值为 536.9μm +/- 27.0(SD),超声为 556.6 +/- 30.5μm,扫描 slit 为 537.2 +/- 32.6μm(声速系数 0.92)。OCT 测量的 CCT 明显比超声薄(P =.007;平均差值-19.7μm;95%可信区间[LoA],-40.4 到 0.9μm),但与扫描 slit 相比无显著差异(P =.2637;平均差值 0.3μm;95% LoA,-24.0 到 23.5μm)。OCT 的 CCT 与超声和扫描 slit 的 CCT 相关性良好(r = 0.940 和 r = 0.934)。
傅里叶域 OCT 角膜厚度测绘具有高度可重复性。与角膜顶点定位相比,瞳孔定位的可重复性更好。超声角膜测厚仪、傅里叶域 OCT 和扫描 slit 体层摄影术不应互换用于 CCT 评估。